Alignment-free characterization of polarizing beamsplitters

Download (5.58 kB)
preprint
posted on 2023-11-30, 18:05 authored by Felipe Calliari, Pedro Tovar, Christiano Nascimento, Breno Perlingeiro, Gustavo Amaral, Guilherme Temporão
Traditional methods for measurement of Polarizing Beamsplitter (PBS) parameters, especially the extinction ratio, require highly polarized light sources, alignment procedures and/or experimental parameters that change over time, such as polarization rotations. In this work, a new method is presented, which employs unpolarized light and a Faraday Mirror. It is shown that precise extinction ratio and insertion loss values can be achieved in three single-sweep measurements, without any alignment requirements or time-varying signals of any kind.