Optica Open
Browse

Brillouin microscopy, what is it really measuring?

Download (5.58 kB)
preprint
posted on 2023-11-30, 17:34 authored by Pei-Jung Wu, Irina Kabakova, Jeffrey Ruberti, Joseph M. Sherwood, Iain E. Dunlop, Carl Paterson, Peter Török, Darryl R. Overby
Brillouin microscopy measures compressibility, but is being increasingly used to assess stiffness of cells and tissues. Using hydrogels with tunable properties, we demonstrate that Brillouin microscopy is insensitive to stiffness of hydrated materials, but depends strongly on water content, consistent with a theoretical model of biphasic compressibility. Empirical correlations between Brillouin measurements and stiffness arise due to their mutual dependence on water content, but correlations vanish once hydration is controlled.

History

Related Materials

Disclaimer

This arXiv metadata record was not reviewed or approved by, nor does it necessarily express or reflect the policies or opinions of, arXiv.