Optica Open
Browse

Extinction spectra of suspensions of microspheres: Determination of spectral refractive index and particle size distribution with nanometer accuracy

Download (5.58 kB)
preprint
posted on 2023-11-30, 17:26 authored by Jonas Gienger, Markus Bär, Jörg Neukammer
A method is presented to infer simultaneously the wavelength-dependent real refractive index (RI) of the material of microspheres and their size distribution from extinction measurements of particle suspensions. To derive the averaged spectral optical extinction cross section of the microspheres from such ensemble measurements, we determined the particle concentration by flow cytometry to an accuracy of typically 2% and adjusted the particle concentration to ensure that perturbations due to multiple scattering are negligible. For analysis of the extinction spectra we employ Mie theory, a series-expansion representation of the refractive index and nonlinear numerical optimization. In contrast to other approaches, our method offers the advantage to simultaneously determine size, size distribution and spectral refractive index of ensembles of microparticles including uncertainty estimation.

History

Disclaimer

This arXiv metadata record was not reviewed or approved by, nor does it necessarily express or reflect the policies or opinions of, arXiv.