High-Resolution Confocal Raman Microscopy Spectrometer Design Based on Echelle Gratings

Download (1.98 MB)
preprint
posted on 2025-11-18, 09:27 authored by jian zhang, xiaotian li, Jiri Jirigalantu, xiaotao mi, yuqi sun, Jian Zhang, geng wang, Guangdong Yu, Fuguan Li, heshig bayan
<p dir="ltr">To address the inherent challenge of balancing broad spectral coverage with high resolution in conventional Raman spectrometers, this paper presents a cross-dispersed confocal Raman spectrometer based on an echelle grating. First, a compound aberration correction system comprising cylindrical and spherical lenses is introduced into the traditional echelle grating spectrometer, effectively suppressing key performance-limiting aberrations such as coma and astigmatism. Compared with traditional echelle grating spectrometers, the resolution and light throughput are improved by approximately a factor of two. Second, a custom confocal Raman collection module enables microscale spot focusing on samples, effectively enhancing Raman signal collection efficiency and improving the signal-to-noise ratio. The designed echelle grating spectrometer and confocal Raman collection module are connected via optical fibers, enabling portability and modularity of the entire system. An echelle grating with a groove density of 52.67 grooves/mm is employed as the main dispersive element, achieving high-resolution detection of 1 cm⁻¹ within a broad wavenumber range of 70–4130 cm⁻¹. Simulation results demonstrate that the aberration correction system reduces the RMS spot radius to below 25 μm, more than doubling the performance compared with conventional designs. Experimental validation confirms the high performance; through calibration with standard light sources and reference materials, high-precision calibration results were obtained. Testing with an Hg lamp demonstrated a full width at half maximum of 1.02 cm⁻¹, closely achieving the design target of 1 cm⁻¹ resolution. Measurements of test substances revealed that the characteristic peak positions and signal-to-noise ratios align with design specifications. This study provides a high-resolution solution for material analysis, biomedical research, and environmental monitoring in laboratory and field settings.</p>

History

Funder Name

National Natural Science Foundation of China (52227810,62205333,U2006209,61975255); Jilin Province Research Projects in China (20230204102YY)

Preprint ID

128881

Usage metrics

    Categories

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC