posted on 2024-02-24, 17:00authored byIsak Silander, Johan Zakrisson, Ove Axner, Martin Zelan
Based on a recent experimental determination of the static polarizability and a first-principles calculation of the frequency-dependent dipole polarizability of argon, this work presents, by use of a Fabry-P\'erot refractometer operated at 1550 nm, a realization of the SI unit of pressure, the pascal, for pressures up to 100 kPa, with an uncertainty of [(0.98 mPa)$^2 + (5.8 \times 10^{-6} P)^2 + (26\times10^{-12}P^2)^2]^{1/2}$. The work also presents a value of the molar polarizability of N$_2$ at 1550 nm of 4.396572(26)$\times 10^{-6}$m$^{3}$/mol, which agrees well with previously determined ones in the literature.
History
Disclaimer
This arXiv metadata record was not reviewed or approved by, nor does it necessarily express or reflect the policies or opinions of, arXiv.